A statistical approach to explore cemented total hip reconstruction performance

Küçük Resim Yok

Tarih

2018

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Springer

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

This study was carried out to determine mechanical behavior and bone adaptation of total hip arthroplasty (THA) subject to concentrated and distributed muscle loads and hip contact forces during activities of walking and stair climbing. Finite element modeling of THA with different prostheses, activity and loading types was developed by adopting a statistical method. Two levels of prostheses, activity, and loading types were selected for the study. 2(3) factorial method was then pursued to design input and output data of finite element analysis. Maximum von Mises stresses were chosen to be output data on which statistical investigation was performed to investigate contribution and interaction of main factors on mechanical failure of cemented THA reconstructions by utilizing analysis of variance method (ANOVA). This study illustrated that the maximum von Mises stresses of THA showed considerable variation for main factors and their two-factor interactions.

Açıklama

Anahtar Kelimeler

Biomechanics, Hip Joint, Finite Element Analysis, Factorial Design, Anova

Kaynak

Australasian Physical & Engineering Sciences In Medicine