Copper sorption efficiency of p-tetranitrocalix[4]arene based high performance liquid chromatographic column
Küçük Resim Yok
Tarih
2016
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Desalination Publ
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Present study deals with the sorption characteristics of p-tetranitrocalix[4]arene (3) appended silica-based high-performance liquid chromatographic (HPLC) column utilized for preconcentration, determination, and purification of Cu2+ from water. The sorption characteristics of the HPLC column were investigated for Cu2+ in column agreement. The experiments were performed in five steps that were monitored using UV-visible diode-array detector. However, all the HPLC experimental results were reconfirmed by using atomic absorption spectrophotometer. The effect of concentration on the sorption efficiency of the column was evaluated for Cu2+ and data obtained was investigated using Langmuir and Freundlich sorption isotherms. The value of coefficient of determination (R-2), i.e. 0.99, suggested that the Freundlich sorption isotherm was found to be the best fit model for Cu2+. The developed method was validated for system efficiency, accuracy, and precision, while the data obtained revealed that the developed column could be applied for the sorption of Cu2+ with adequate efficiency and repeatability.
Açıklama
Anahtar Kelimeler
Hplc, Aas, Copper, Sorption, Calixarene
Kaynak
Desalination And Water Treatment
WoS Q Değeri
Q2
Scopus Q Değeri
Q3
Cilt
57
Sayı
19