Comparison of cone-beam computed tomography with bitewing radiography for detection of periodontal bone loss and assessment of effects of different voxel resolutions: an in vitro study

Küçük Resim Yok

Tarih

2019

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Springer

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

ObjectiveThe aim of the study was to compare intraoral radiographs and CBCT images for detection of horizontal periodontal bone loss, and to investigate the diagnostic effect of different voxel resolutions in CBCT imaging.MethodsA total of 240 sites with horizontal bone loss were measured on the buccal, lingual, mesial, and distal surfaces of 60 posterior teeth in four maxillary and six mandibular bones obtained from cadavers (dry skulls). Direct measurements on the dry skulls were accepted as the gold standard values. Measurements on CBCT images at two different voxel resolutions (0.250 and 0.160 mm(3)) and intraoral bitewing radiographs were compared with one another and with the gold standard values.ResultsThe measurements on the CBCT images at two voxel resolutions and bitewing radiographs did not differ significantly (p>0.05) from the direct measurements on the dry skulls. No significant difference was found between the bitewing radiographs and CBCT images for measurements in the mesial and distal regions (p>0.05). There was no significant difference between the measurements on the buccal and lingual surfaces at the two different voxel resolutions (p>0.05).ConclusionsCBCT scans are recommended for evaluation of buccal and lingual bone loss to avoid intraoral radiographs that exceed routine examination of interproximal alveolar bone loss. Furthermore, instead of basing the voxel size on the required CBCT scans, it is recommended to select the smallest possible field of view to reduce the dose of radiation.

Açıklama

Anahtar Kelimeler

Cbct, Voxel, Periodontal Bone Loss, Bitewing Radiography

Kaynak

Oral Radiology

WoS Q Değeri

Q4

Scopus Q Değeri

Q2

Cilt

35

Sayı

2

Künye